Bio Safety Cabinet

JisicoJ-BSCV1, J-BSCV2

1

● Vertical type is used for jobs where ventilation of gases generated out of the experiments is required.
● Dust collection is carried out by HEPA filter (main filter).
● Approximately 70% of the HEPA-filtered air is circulated through the cabinet, while 30% passes through an exhaust HEPA filter and is
discharged.
● For vertical type, various instruments and equipment can be placed on the working-bench, for the air flow with negative pressure.
● Exact life span of filter is displayed in LCD display by adopting air flow sensor mode.
● Facility for working : A low-noise fan is attached in it and it is easy for working on account of flatten side of stay for working.
● Various facilities : It sterilizes the working space with ultraviolet lamp.

Model

J-BSCV1

J-BSCV2

Type

Class II A2

Overall size(W×D×H)

mm

 1,050×785×2,230/2,280

1,380×785×2,230/2,280

Inner size(W×D×H)

mm

900×645×730

1,230×645×730

Working size(W×D)

mm

860×475

1,190×475

Stand size(W×D×H)

mm

800×720×670/710

1,280×720×670/710

Main control

LCD microprocessor control with membrane touch type

Main filter

HEPA filter(99.97% efficient on particle of 0.3μ)
Option:ULPA filter(99.999% efficient on particle of 0.3μ)

FL lamp

W

20×1ea

40×1ea

Germicidal lamp(UV)

W

30×1ea

Air flowing volume

/min

45±5

Velocity

m/sec

0.5 ± 0.2

Illumination

Lux

More than 1,200

Blower motor

Hp

1/2

Door

Clear safety reinforced glass, 5mm thickness

Utility

socket

AC 220V 2ea

gas cock

1ea

Stand caster

Foot master

Material

working table

Stainless steel plate

cabinet

Steel plate with powder coated finish

Power source

AC 230V, 50/60Hz, 1phase

AC 230V, 50/60Hz, 1phase

Certification

CE

Share this post