SX-Series

Tomy

1

Saving the installation space

The main body is very compact owing to the design which provides thorough installation space saving. In addition, installation space at the side of the main body which is required for models employing a slide-open lid, which is not necessary at SX autoclaves top-open lid. Even when several SX autoclaves are installed side-by-side, the necessary installation space is minimized.

A : 790mm(SX-300),970mm(SX-500),1003mm(SX-700)
B : 614mm(SX-300),794mm(SX-500),827mm(SX-700)
C : 410mm(SX-300/500),470mm(SX-700)

2

Easily viewable “work monitor”

The LED display shows working status for easy monitoring. The process being performed is indicated by an LED indication lamp blinking in red.

The illuminated display on the operating panel shown in the picture is different from the actual display at the time of operation.

Cooling fan provided as a standard feature

A rapid air-cooling function (vessel-cooling fan) permitting reduction of chamber cool down time is provided as a standard feature. This is most appropriate for lowering the temperature rapidly on completion of the cycle. The time required for lower the temperature is much reduced by employment of the cooling fan in comparison with natural cooling.

Sterilizing course selection

The optimal sterilizing course can be selected from among five courses, such as the liquid sterilizing course in addition to that for normal sterilization.

3

Other features

Large indication lamp “Operated”

An easily viewable large indication “Operated” is provided at the upper section of the operating panel.
Depending on the pressure status (normal pressure/pressure is applied),the displayed color changes.

Setting to 100°C is possible.

Variable temperature setting such as to 100°C as well as to 121°C has been made possible.The temperature can be set within the range from 45°C to 135°C in steps of one degree (heating mode up to 104°C).

Auto-variable exhaust speed

A function allowing the exhaust valve to open automatically after completion of sterilization is provided. The exhaust speed can be set to one of six levels (set to variable for liquid sterilization)

Timer function

The operation starting time can be preset easily. By presetting the start of operation, otherwise wasted periods at night or early morning, can be used effectively.

Setting to 100°C is possible.4

Variable temperature setting such as to 100°C as well as to 121°C has been made possible.The temperature can be set within the range from 45°C to 135°C in steps of one degree (heating mode up to 104°C).

Lid interlock

A safety device (lid interlock mechanism) for locking the lid during operation is provided. A high level of safety is assured to prevent unanticipated accidents.

Pressure fine adjustment function

When the balance between the temperature and pressure deviates during sterilization, fine exhausting is carried out automatically in order to adjust the best chamber condition.

Functions & Features
Liquid sterilizing / Heating / Warming / Check / Last-run memory / Built-in steam exhaust bottle
Flat internal surface of the chamber / Lid opening/closing detection mechanism
Exhaust bottle detection mechanism / Leakage breaker provided

OPTION

External sensor for the articles to be sterilized

5

 

Sensor for directly detecting the temperature of articles to be sterilized. More reliable sterilization monitor can be carried out.

Data output

6

The temperature data, pressure data, etc. can be exported. The temperature can be recorded when the external recorder is connected.

External sensor and recorder

7

A temperature sensor and recorder independent of the autoclaves. The chamber temperature can be recorded.


SX Specifications

Model name
SX-300
SX-500
SX-700
Operating temperature range
Sterilizing
105 ∼ 135°C(0.019 ∼ 0.212MPa)
Heating
45 ∼ 104°C(0 ∼ 0.015MPa)
Warming
45 ∼ 95°C
Maximum operating pressure
0.263MPa
0.25MPa
Temperature
Display
Digital
Pressure gauge
Display
Analog
Display range
0 ∼ 0.4MPa
Heat source
1.5kW electric heater
2kW electric heater
3kW electric heater
Safety device
• Water level sensor, • Current leakage breaker, • Lid interlock, • Over-heat prevention, • Over-pressure prevention, • Open temperature sensor detection, • Safety valve
Time
Display
Digital
Display
range
Sterilizing
0:00 – 9:59, 0 – 99hours, 0 – 999min.(0:00 – 9:59,10 – 99hours changeable)
Heating
Warming
Variable: 0 – 99hours
Chamber dimensions
φ325 x 553mm
φ325 x 733mm
φ370 x 774mm
Chamber capacity
Effective internal
volume:36L,
Internal volume:44L
Effective internal
volume:50L,
Internal volume:58L
Effective internal
volume:69L,
Internal volume:79L

Chamber material

SUS304

Dimensions (mm)
410W x 477D x 790Hmm
(With protruding:574D)
410W x 477D x 970Hmm
(With protruding:574D)
470W x 528D x 1003Hmm
(With protruding:625D)
Weight
50kg
60kg
72kg
Rated voltage
120V AC
230V AC
Power input
13A/120V
17A/120V
7A/230V
9A/230V
13A/230V
Required power supply
Single-phase 120V AC
(50/60 Hz)15A or more
Single-phase 120V AC
(50/60 Hz)20A or more
Single-phase 230V AC
(50/60 Hz)10A or more
Single-phase 230V AC
(50/60 Hz)15A or more
Single-phase 230V AC
(50/60 Hz)15 A or more
Power consumption (calorific value)
1.5kW (1290kcal/h)
2.0kW (1720cal/h)
3.0kW (2580cal/h)
Environmental Conditions
When opereting the autoclave, observe the environmental conditions given below.
Ambient temperature: 10 to 35°C
Almospheric pressure: 860 to 1060hPa
Relative humidity: 30 to 85%
Maximum gradient: 2
Accessories
Stainless baskets 1
(φ300 x 182mm)
Stainless baskets 2
(φ300 x 182mm)
Stainless baskets 2
(φ345 x 181mm)
• Chanber bottom Plate 1, • Caster stoppers 4, • Operator’s Manual 1, • Clear folder (for storing the operation manual) 1, • Screw (for attaching the clear folder) 1, • Warranty card 1, • Customer card 1, • Inspection Sheet 1